Packaging for terminally sterilized medical devices - Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) - Requirements and test methods

This part of EN 868 provides test methods and values for paper, used in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) used as sterile barrier systems and/or packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use.
NOTE 1   The need for a protective packaging may be determined by the manufacturer and the user.
This part of EN 868 only introduces performance requirements and test methods that are specific to the products covered by this part of EN 868 but does not add or modify the general requirements specified in EN ISO 11607-1.
As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more but not all of the requirements of EN ISO 11607-1.
NOTE 2   When additional materials are used inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, container filter indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) then other requirements, including the determination of the acceptability of these materials during validation activities, may apply.
The materials specified in this part of EN 868 are intended for single use only.
NOTE 3   Applicable sterilization methods are specified by the manufacturer.

Verpackungsmaterialien für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 3: Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-5) - Anforderungen und Prüfverfahren

Dieser Teil der EN 868 enthält Prüfverfahren und Werte für Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868 4) und zur Herstellung von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868 5).
Dieser Teil der EN 868 enthält keine zusätzlichen Anforderungen zu den in EN ISO 11607-1 enthaltenen allgemeinen Anforderungen.
Die besonderen Anforderungen in 4.2 können verwendet werden, um Übereinstimmung mit einer oder mehreren Anforderungen – aber nicht mit allen Anforderungen  der EN ISO 11607-1 nachzuweisen.
Das in diesem Teil festgelegte Papier ist zur Verwendung als Verpackungsmaterial für Medizinprodukte geeignet, die in ihrer Endverpackung sterilisiert werden sollen.
ANMERKUNG   Anwendbare Sterilisationsverfahren werden durch den Hersteller festgelegt.

Matériaux et systèmes d'emballage pour les dispositifs médicaux stérilisés au stade terminal - Partie 3: Papier utilisé dans la fabrication des sacs en papier (spécifiés dans l'EN 868-4) et dans la fabrication de sachets et gaines (spécifiés dans l'EN 868-5) - Exigences et méthodes d'essai

La présente partie de l’EN 868 fournit tant méthodes d’essai que valeurs relatives au papier utilisé dans la
fabrication des sacs en papier (spécifiés dans l’EN 868-4) et dans la fabrication de sachets et gaines
(spécifiés dans l’EN 868-5) utilisés comme systèmes de barrière stérile et/ou comme systèmes d’emballage
destinés à maintenir l’état de stérilité des dispositifs médicaux, stérilisés au stade terminal, jusqu’à leur
utilisation.
NOTE 1 La nécessité d’un emballage de protection peut être déterminée par le fabricant et l’utilisateur.
La présente partie de l’EN 868 présente uniquement les exigences de performance et les méthodes d’essai
spécifiques des produits couverts par la présente partie de l’EN 868 mais elle n'ajoute pas ou ne modifie pas
les exigences générales spécifiées dans l’EN ISO 11607-1.
En conséquence, les exigences particulières indiquées dans le paragraphe 4.2 peuvent être utilisées pour
démontrer la conformité à une ou plusieurs, mais pas à toutes les exigences de l’EN ISO 11607-1.
NOTE 2 En cas d'utilisation de matériaux supplémentaires à l’intérieur du système de barrière stérile de manière à
faciliter le rangement, le séchage ou la présentation aseptique (par exemple, enveloppe interne, filtre de conteneur,
indicateurs, listes d’emballage, tapis, dispositifs de rangement des instruments, fonds de paniers ou une enveloppe
supplémentaire entourant le dispositif médical), d’autres exigences peuvent alors s’appliquer, y compris la détermination
de l’acceptabilité de ces matériaux lors des activités de validation.
Les matériaux spécifiés dans la présente partie de l’EN 868 sont destinés exclusivement à un usage unique.
NOTE 3 Les méthodes de stérilisation applicables sont spécifiées par le fabricant.

Embalaža za končno sterilizirane medicinske pripomočke - 3. del: Papir za izdelavo papirnatih vrečk (specifikacija EN 868-4) in papir za izdelavo vrečk in neskončnih zvitkov (specifikacija EN 868-5) - Zahteve in preskusne metode

General Information

Status
Withdrawn
Publication Date
09-Sep-2009
Withdrawal Date
27-Feb-2017
Technical Committee
Current Stage
9900 - Withdrawal (Adopted Project)
Start Date
27-Feb-2017
Due Date
22-Mar-2017
Completion Date
28-Feb-2017

Relations

Buy Standard

Standard
EN 868-3:2009
English language
16 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Verpackungsmaterialien für in der Endverpackung zu sterilisierende Medizinprodukte - Teil 3: Papier zur Herstellung von Papierbeuteln (festgelegt in EN 868-4) und zur Herstellung von Klarsichtbeuteln und -schläuchen (festgelegt in EN 868-5) - Anforderungen und PrüfverfahrenMatériaux et systèmes d'emballage pour les dispositifs médicaux stérilisés au stade terminal - Partie 3: Papier utilisé dans la fabrication des
sacs en papier (spécifiés dans l'EN 868-4) et dans la fabrication de sachets et gaines (spécifiés dans l'EN 868-5) - Exigences et méthodes d'essaiPackaging for terminally sterilized medical devices - Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) - Requirements and test methods11.080.30Sterilizirana embalažaSterilized packagingICS:Ta slovenski standard je istoveten z:EN 868-3:2009SIST EN 868-3:2009en,fr,de01-oktober-2009SIST EN 868-3:2009SLOVENSKI
STANDARDSIST EN 868-3:20001DGRPHãþD



SIST EN 868-3:2009



EUROPEAN STANDARDNORME EUROPÉENNEEUROPÄISCHE NORMEN 868-3May 2009ICS 11.080.30Supersedes EN 868-3:1999
English VersionPackaging for terminally sterilized medical devices - Part 3:Paper for use in the manufacture of paper bags (specified in EN868-4) and in the manufacture of pouches and reels (specified inEN 868-5) - Requirements and test methodsMatériaux d'emballage pour les dispositifs médicauxstérilisés au stade terminal - Partie 3: Papier utilisé dans lafabrication de sacs en papier (spécifiés dans l'EN 868-4) etdans la fabrication de sachets et gaines (spécifiés dansl'EN 868-5) - Exigences et méthodes d'essaiVerpackungen für in der Endverpackung zu sterilisierendeMedizinprodukte - Teil 3: Papier zur Herstellung vonPapierbeuteln (festgelegt in EN 868-4) und zur Herstellungvon Klarsichtbeuteln und -schläuchen (festgelegt in EN868-5) - Anforderungen und PrüfverfahrenThis European Standard was approved by CEN on 23 April 2009.CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this EuropeanStandard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such nationalstandards may be obtained on application to the CEN Management Centre or to any CEN member.This European Standard exists in three official versions (English, French, German). A version in any other language made by translationunder the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as theofficial versions.CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland,France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal,Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.EUROPEAN COMMITTEE FOR STANDARDIZATIONCOMITÉ EUROPÉEN DE NORMALISATIONEUROPÄISCHES KOMITEE FÜR NORMUNGManagement Centre:
Avenue Marnix 17,
B-1000 Brussels© 2009 CENAll rights of exploitation in any form and by any means reservedworldwide for CEN national Members.Ref. No. EN 868-3:2009: ESIST EN 868-3:2009



EN 868-3:2009 (E) 2 Contents Page Foreword .3Introduction .41Scope .52Normative references .53Terms and definitions .64Requirements .65Information to be supplied by the manufacturer .7Annex A (informative)
Details of significant technical changes between this European Standard and the previous edition .8Annex B (normative)
Method for the determination of fluorescence .9Annex C (normative)
Method for the determination of water repellency . 10Annex D (normative)
Method for the determination of pore size . 11Bibliography . 16 SIST EN 868-3:2009



EN 868-3:2009 (E) 3 Foreword This document (EN 868-3:2009) has been prepared by Technical Committee CEN/TC 102 “Sterilizers for medical purposes”, the secretariat of which is held by DIN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by November 2009, and conflicting national standards shall be withdrawn at the latest by November 2009. Attention is drawn to the possibility that some of the elements of this document may be the subject of patent rights. CEN [and/or CENELEC] shall not be held responsible for identifying any or all such patent rights. This document supersedes EN 868-3:1999. Annex A provides details of significant technical changes between this European Standard and the previous edition. EN 868 consists of the following parts, under the general title Packaging for terminally sterilized medical devices:
Part 2: Sterilization wrap — Requirements and test methods; Part 3: Paper for use in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) — Requirements and test methods; Part 4: Paper bags — Requirements and test methods; Part 5: Sealable pouches and reels of porous materials and plastic film construction — Requirements and test methods; Part 6: Paper for low temperature sterilization processes — Requirements and test methods; Part 7: Adhesive coated paper for low temperature sterilization processes — Requirements and test methods; Part 8: Re-usable sterilization containers for steam sterilizers conforming to EN 285 — Requirements and test methods; Part 9: Uncoated nonwoven materials of polyolefines — Requirements and test methods; Part 10: Adhesive coated nonwoven materials of polyolefines — Requirements and test methods. In addition, ISO/TC 198 "Sterilization of health care products" in collaboration with CEN/TC 102 "Sterilizers for medical purposes" has prepared the EN ISO 11607 series "Packaging for terminally sterilized medical devices". The EN ISO 11607 series specifies general requirements for materials, sterile barrier systems and packaging systems (Part 1) and validation requirements for forming, sealing and assembly processes (Part 2). According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and the United Kingdom. SIST EN 868-3:2009



EN 868-3:2009 (E) 4 Introduction The EN ISO 11607 series consists of two parts under the general title "Packaging for terminally sterilized medical devices". Part 1 of this series specifies general requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. Part 2 of this series specifies validation requirements for forming, sealing and assembly processes.
Every sterile barrier system shall fulfil the requirements of EN ISO 11607-1. The EN 868 series can be used to demonstrate compliance with one or more of the requirements specified in EN ISO 11607-1.
During the revision of EN 868 parts 2 to 10 CEN/TC 102/WG 4 recognized Resolution CEN/BT 21/2003 relating to the implementation of the uncertainty of measurement concept in standards. Following this Resolution and the corresponding guidance, CEN/TC 102/WG 4 has initiated a review of the test methods needed to show compliance with the requirements specified in EN 868 parts 2 to 10 with the intention that the information required by CEN/BT 21/2003 be available for inclusion in EN 868 parts 2 to 10 during one of their next revisions.
CEN/TC 102/WG 4 also appreciates the initiatives of CEN with regard to the minimization of adverse environmental impacts by standards. It was agreed that this subject should be given priority during the next edition of the EN ISO 11607 series that is the basic reference for all parts of the EN 868 series.
SIST EN 868-3:2009



EN 868-3:2009 (E) 5 1 Scope This part of EN 868 provides test methods and values for paper, used in the manufacture of paper bags (specified in EN 868-4) and in the manufacture of pouches and reels (specified in EN 868-5) used as sterile barrier systems and/or packaging systems that are intended to maintain sterility of terminally sterilized medical devices to the point of use. NOTE 1 The need for a protective packaging may be determined by the manufacturer and the user.
This part of EN 868 only introduces performance requirements and test methods that are specific to the products covered by this part of EN 868 but does not add or modify the general requirements specified in EN ISO 11607-1. As such, the particular requirements in 4.2 can be used to demonstrate compliance with one or more but not all of the requirements of EN ISO 11607-1. NOTE 2 When additional materials are used inside the sterile barrier system in order to ease the organization, drying or aseptic presentation (e.g. inner wrap, container filter indicators, packing lists, mats, instrument organizer sets, tray liners or an additional envelope around the medical device) then other requirements, including the determination of the acceptability of these materials during validation activities, may apply.
The materials specified in this part of EN 868 are intended for single use only. NOTE 3 Applicable sterilization methods are specified by the manufacturer. 2 Normative references The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies. EN 20187, Paper, board and pulps — Standard atmosphere for conditioning and testing and procedure for monitoring the atmosphere and conditioning of samples (ISO 187:1990) EN 20535, Paper and board — Determination of water absorptiveness — Cobb method (ISO 535:1991) EN 21974, Paper — Determination of tearing resistance (Elmendorf method) (ISO 1974:1990) EN ISO 536, Paper and board — Determination of grammage (ISO 536:1995) EN ISO 1924-2, Paper and board — Determination of tensile properties — Part 2: Constant rate of elongation method (ISO 1924-2:1994) EN ISO 2758, Paper — Determination of bursting strength (ISO 2758:2001) EN ISO 11607-1:2006, Packaging for terminally sterilized medical devices — Part 1: Requirements for materials, sterile barrier systems and packaging systems (ISO 11607-1:2006) ISO 3689, Paper and board — Determination of bursting strength after immersion in water ISO 3781, Paper and board — Determination of tensile strength after immersion in water ISO 5636-3, Paper and board — Determination of air permeance (medium range) — Part 3: Bendtsen method ISO 6588-2:2005, Paper, board and pulps — Determination of pH of aqueous extracts — Part 2: Hot extraction SIST EN 868-3:2009



EN 868-3:2009 (E) 6 ISO 8601, Data elements and interchange formats — Information interchange — Representation of dates and times
ISO 9197, Paper, board and pulps — Determination of water-soluble chlorides ISO 9198, Paper, board and pulp — Determination of water-soluble sulfates 3 Terms and definitions For the purposes of this document, the terms and definitions given in EN ISO 11607-1:2006 apply. 4 Requirements 4.1 General The requirements of EN ISO 11607-1 apply. NOTE EN ISO 11607-1:2006, 5.1.4 refers to conditions during production and handling with respect to their impact on the product (e.g. electrostatic conductivity, bioburden if applicable).
4.2 Performance requirements and test methods 4.2.1 The requirements of 4.2.11 and 4.2.15 do not apply to materials solely used in irradiation sterilization packaging. 4.2.2 No colour shall leach out of the paper. Compliance shall be tested by visual examination of a hot aqueous extract prepared in accordance with the method given in ISO 6588-2. 4.2.3 The average mass of 1 m² of the conditioned paper when tested in accordance with EN ISO 536 shall be within ± 5 % of the nominal value stated by the manufacturer. 4.2.4 The pH of an aqueous extract of the paper shall be not less than 5 nor greater than 8 when tested in accordance with ISO 6588-2.
4.2.5 The chloride content of the paper, calculated as sodium chloride, shall not exceed 0,05 % when tested in accordance with ISO 9197 using a hot extract prepared in accordance with ISO 6588-2:2005, 7.2 except that 2 ml of potassium chloride solution is not added.
4.2.6 The sulphate content of the paper, calculated as sodium sulphate, shall not exceed 0,25 % when tested in accordance with ISO 9198, using a hot extract prepared in accordance with ISO 6588-2:2005, 7.2 except that 2 ml of potassium chloride solution is not added.
4.2.7 When tested in accordance with Annex B the paper shall neither exhibit an increase in brightness due to the optical brightener of more than 1 % nor have more than five fluorescent spots, each having an axis greater than 1 mm per 0,01 m². 4.2.8 The internal tearing resistance of the conditioned paper shall be not less than 550 mN in both machine and cross direction when tested in accordance with EN 21974. 4.2.9 The air permeance of the conditioned paper shall be not less than 3,4 µm/Pa ⋅ s at an air pressure of 1,47 kPa when tested in accordance with ISO 5636-3. 4.2.10 The bursting strength of the conditioned paper shall be not less than 230 kPa when tested in accordance with EN ISO 2758. SIST EN 868-3:2009



EN 868-3:2009 (E) 7 4.2.11 The wet bursting strength of the paper shall be not less than 35 kPa when tested in accordance with ISO 3689 using an immersion time of 10 min. 4.2.12 The water repellency of the paper shall be such that the penetration time is not less than 20 s when tested in accordance with Annex C. 4.2.13 When tested in accordance with Annex D, the average of the pore diameters of the ten test pieces shall be lower than or equal to 35 µm. No value shall be greater than 50 µm. 4.2.14 The tensile strength of the conditioned paper shall be not less than 4,40 kN/m in machine direction and not less than 2,20 kN/m in cross direction when tested in accordance with EN ISO 1924-2. 4.2.15 The wet tensile strength of the paper shall be not less than 0,90 kN/m in
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.