Method for analysis of allergens - Quantification of an extended list of 57 suspected allergens in ready to inject fragrance materials by gas chromatography mass spectrometry

The proposed standard aims at describing a method to analyze 57 chemically defined suspected allergens (some of them existing under several isomeric forms or as mixtures) in ready to inject fragrance and raw material samples according to the SCCS opinion. (SCCS/1459/11).
This new analytical method uses gas chromatography and mass spectrometry (GC-MS) to detect and to quantify the 57 fragrance substances and their relevant isomers at a concentration higher than 0.0002% (2 mg/kg) in ready to inject fragrance and raw material samples. Making this method available will allow the screening of (complex) ready to inject fragrance and raw material samples to be undertaken for the presence of any of those chemically defined suspected allergens. It will therefore be a basis for the calculation of adequate information to the cosmetics industry in order to provide adequate consumer information. The present analytical method uses GC-MS by combination of two GC columns of different polarity with a dedicated methodology for quantitation.

Analyseverfahren für Allergene - Quantifizierung einer erweiterten Liste von 57 zu vermutenden Allergenen in einspritzfertigen Duftstoffen mittels Gaschromatographie/Massenspektrometrie

Das vorliegende Verfahren ermöglicht die Identifizierung und Quantifizierung der als Allergene vermuteten flüchtigen Verbindungen, die in den in kosmetischen Mitteln verwendeten Verbindungen von Duftstoffen und Rohstoffen von Duftstoffen enthalten sind. Die Analyse wird mithilfe der Gaschromatographie und Massenspektrometrie (GC MS) an Matrixproben durchgeführt, die „einspritzfertig“ und mit der Gaschromatographie kompatibel sind.
Die von diesem Verfahren erfassten Analyten basieren auf den Inhalten der in der Stellungnahme des SCCS 1459/11 (1) enthaltenen Tabelle 13.1 und Tabelle 13.2 und den von der Europäischen Kommission vorgeschlagenen aufgeführten Rechtsvorschriften. Die Begründung für die endgültige Auswahl der während des Verfahrens angewendeten Analyten ist in der in Anhang J enthaltenen Tabelle aufgeführt.
Das Verfahren wurde auf IFRA  und CEN Ebene validiert.

Méthode d’analyse des allergènes - Quantification de la liste étendue des 57 allergènes suspectés dans les matières premières de parfumerie et les compositions parfumantes prêtes à être injectées, par chromatographie en phase gazeuse/spectrométrie de masse

La présente méthode permet l’identification et la quantification des composés volatils suspectés allergènes, qui sont présents dans les matières premières de parfumerie et les compositions parfumantes utilisées dans les produits cosmétiques. L’analyse est réalisée par chromatographie en phase gazeuse couplée à la spectrométrie de masse (CG-SM) sur des échantillons de matrice qui sont « prêts à être injectés » et qui sont compatibles avec la chromatographie en phase gazeuse.
Les analytes couverts par ce mode opératoire sont basés sur le contenu des Tableaux 13.1 et 13.2 du document d’opinion CSSC 1459/11 (1) et sont mentionnés dans la législation proposée par la Commission européenne. Les raisons qui justifient le choix final des analytes pour ce mode opératoire sont données dans le tableau figurant à l’Annexe J.
La méthode a été validée au niveau de l’IFRA et du CEN.

Analizne metode za alergene - Kvantitativno določevanje 57 domnevnih alergenov z razširjenega seznama v dišavnih izdelkih s plinsko kromatografsko analizo vzorcev, ki so pripravljeni za injeciranje, in masno spektrometrijo

Namen predlaganega standarda je opisati metodo za analizo 57 kemično določenih domnevnih alergenov (nekateri od njih obstajajo v več izomernih oblikah ali kot mešanice) v vzorcih dišav in surovin, ki so pripravljeni za injiciranje v skladu z mnenjem SCCS (SCCS/1459/11).
Ta nova analitična metoda uporablja plinsko kromatografijo in masno spektrometrijo (GC-MS) za odkrivanje in kvantitativno določevanje 57 dišavnih snovi in njihovih pomembnih izomerov s koncentracijo, ki presega 0,0002 % (2 mg/kg), v vzorcih dišav in surovin, ki so pripravljeni za injiciranje. Z zagotavljanjem te metode bo omogočeno presejanje (kompleksnih) vzorcev dišav in surovin, ki so pripravljeni za injiciranje, s čimer se ugotavlja prisotnost katerega koli od teh kemično določenih domnevnih alergenov. To bo torej osnova za izračun ustreznih informacij, na podlagi katerih bo lahko kozmetična industrija zagotovila ustrezno obveščanje potrošnikov. Ta analitična metoda uporablja plinsko kromatografijo in masno spektrometrijo s kombinacijo dveh kolon GC različne polarnosti s posebno metodologijo za kvantitativno določevanje.

General Information

Status
Published
Public Enquiry End Date
02-Apr-2020
Publication Date
12-Dec-2021
Current Stage
6060 - National Implementation/Publication (Adopted Project)
Start Date
12-Aug-2021
Due Date
17-Oct-2021
Completion Date
13-Dec-2021

Relations

Buy Standard

Standard
EN 16274:2022 - BARVE
English language
70 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day
Draft
prEN 16274:2020 - BARVE
English language
69 pages
sale 10% off
Preview
sale 10% off
Preview
e-Library read for
1 day

Standards Content (Sample)

SLOVENSKI STANDARD
SIST EN 16274:2022
01-januar-2022
Nadomešča:
SIST EN 16274:2012
Analizne metode za alergene - Kvantitativno določevanje 57 domnevnih alergenov
z razširjenega seznama v dišavnih izdelkih s plinsko kromatografsko analizo
vzorcev, ki so pripravljeni za injeciranje, in masno spektrometrijo
Method for analysis of allergens - Quantification of an extended list of 57 suspected
allergens in ready to inject fragrance materials by gas chromatography mass
spectrometry
Analyseverfahren für Allergene - Quantifizierung einer erweiterten Liste von 57 zu
vermutenden Allergenen in einspritzfertigen Duftstoffen mittels
Gaschromatographie/Massenspektrometrie
Méthode d’analyse des allergènes - Quantification de la liste étendue des 57 allergènes
suspectés dans les matières premières de parfumerie et les compositions parfumantes
prêtes à être injectées, par chromatographie en phase gazeuse/spectrométrie de masse
Ta slovenski standard je istoveten z: EN 16274:2021
ICS:
71.040.50 Fizikalnokemijske analitske Physicochemical methods of
metode analysis
71.100.70 Kozmetika. Toaletni Cosmetics. Toiletries
pripomočki
SIST EN 16274:2022 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
SIST EN 16274:2022

---------------------- Page: 2 ----------------------
SIST EN 16274:2022


EN 16274
EUROPEAN STANDARD

NORME EUROPÉENNE

May 2021
EUROPÄISCHE NORM
ICS 71.100.60 Supersedes EN 16274:2012
English Version

Method for analysis of allergens - Quantification of an
extended list of 57 suspected allergens in ready to inject
fragrance materials by gas chromatography mass
spectrometry
Méthode d'analyse des allergènes - Quantification de la Analyseverfahren für Allergene - Quantifizierung einer
liste étendue des 57 allergènes suspectés dans les erweiterten Liste von 57 zu vermutenden Allergenen
matières premières de parfumerie et les compositions in einspritzfertigen Duftstoffen mittels
parfumantes prêtes à être injectées, par Gaschromatographie/Massenspektrometrie
chromatographie en phase gazeuse/spectrométrie de
masse
This European Standard was approved by CEN on 8 February 2021.

CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references
concerning such national standards may be obtained on application to the CEN-CENELEC Management Centre or to any CEN
member.

This European Standard exists in three official versions (English, French, German). A version in any other language made by
translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC Management
Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.





EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CEN All rights of exploitation in any form and by any means reserved Ref. No. EN 16274:2021 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
Contents                                                         Page
European foreword . 4
Introduction . 5
1 Scope . 6
2 Normative references . 6
3 Terms and definitions . 6
4 Principle . 6
5 Reagents . 7
5.1 Solvents . 7
5.1.1 Methyl pivalate . 7
5.1.2 Tert-Butyl Methyl Ether (MTBE) . 7
5.2 Reference samples (suspected allergens) . 7
5.2.1 General. 7
6 Apparatus . 11
6.1 Gas chromatograph equipped with flame ionization detector (GC-FID) . 11
6.2 Gas chromatograph coupled to a mass spectrometer (GC-MS) . 11
6.2.1 General. 11
6.2.2 GC-MS system. 11
6.3 Capillary columns for GC . 11
6.4 Analytical balance . 12
7 Stock and sample solutions - preparation and storage . 12
7.1 General information . 12
7.1.1 Choice of the solvent . 12
7.1.2 Miscellaneous . 13
7.2 Preparation of stock solutions from reference samples . 13
7.3 Preparation of internal standard solutions and calibration solutions . 13
7.3.1 General. 13
7.3.2 Stock Solutions of Internal Standards (1g/kg equivalent) . 13
7.3.3 Calibration Solutions Concentrations . 13
7.3.4 Preparation of Calibration solutions . 14
8 Mass spectrometer acquisition conditions . 16
8.1 Establishment of retention times and SIM chromatograph windows . 16
8.2 SIM Window – Set up and criteria . 17
8.3 GC-MS scan mode verification . 17
9 Sample analysis . 17
9.1 General. 17
9.2 Sample preparation for analysis . 17
9.3 Sequence . 18
9.3.1 Blank . 18
9.3.2 Calibration solutions . 18
9.3.3 Check solution . 18
9.3.4 Samples . 19
10 Data validation and treatment . 19
10.1 General. 19
10.2 Examination of Q values . 19
10.3 Relative intensity of characteristic ions in SIM . 20
2

---------------------- Page: 4 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
10.4 Relative ion intensity using scan data . 20
10.5 Data verification scheme and reporting of final concentration . 21
10.6 GC-MS scan mode verification . 21
10.7 Mass spectrum evaluation in SCAN mode . 21
11 Test report . 21
Annex A (informative) Determination of calibrant and reference sample purity . 22
Annex B (informative) GC capillary column parameters . 29
Annex C (informative) SIM Ions for SIM or SIM-SCAN Use. 30
Annex D (informative) Example of SIM windows . 37
Annex E (informative) Example of chromatograms . 46
Annex F (normative) Decisional tree for quantification of suspected allergens . 48
Annex G (informative) Preparation of stock solutions from reference samples . 49
Annex H (informative) Calibration methods and approach . 54
Annex I (informative) Quantification of allergens . 56
Annex J (informative) Suspected allergens selected as analytes targets – Rationale . 59
Annex K (informative) Other general information. 69
Bibliography . 70

3

---------------------- Page: 5 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
European foreword
This document (EN 16274:2021) has been prepared by Technical Committee CEN/TC 347 “Methods for
analysis of allergens”, the secretariat of which is held by SNV.
This European Standard shall be given the status of a national standard, either by publication of an
identical text or by endorsement, at the latest by November 2021, and conflicting national standards shall
be withdrawn at the latest by November 2021.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CEN shall not be held responsible for identifying any or all such patent rights.
This document supersedes EN 16274:2012.
The most significant technical changes made in comparison to the previous version are as follows:
— Number of allergens has been extended from 26 to 57 chemically defined molecules.
— Solvents: moves from methyl pivalate and ortho fluorotoluene and acetone to methyl pivalate and
MTBE and other solvents provided they are tested prior being used.
— Sample preparation: allows to use weight/volume approach rather than only weight/weight.
— Data processing: provides explanations for a calibration curve using a forced through zero approach.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directive(s).
According to the CEN-CENELEC Internal Regulations, the national standards organisations of the
following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Croatia,
Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland,
Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Republic of North
Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and the United
Kingdom.
4

---------------------- Page: 6 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
Introduction
Directive 2003/15/EC amending Council Directive 76/768/EEC, relating to cosmetic products, regulates
the obligation to inform consumers of the presence of 24 chemically-defined fragrance substances
identified as potential allergens in cosmetic products. Following the publication of the Scientific
Committee on Consumer Safety’s document (SCCS/1459/11), it was proposed to extend that to 57
fragrance substances, some of them existing under several isomeric forms or as mixtures. This required
the development of a new quantification method in response to the evolution of regulatory requirements.
The new analytical method has been developed using gas chromatography and mass spectrometry (GC-
MS), to detect and to quantify the 57 fragrance substances and their relevant isomers in ready to inject
fragrance compounds and fragrance raw materials.
The method described in this document does not include requirements for the preparation of samples in
matrices for which direct injection in GC is not feasible.
The present document describes a working analytical method based on IFRA Analytical Working Group
developments. The analytical method was validated based on a ring test performed by the CEN working
group using the accuracy profile approach.

5

---------------------- Page: 7 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
1 Scope
The present method permits the identification and quantification of the volatile compounds suspected as
allergens, which are present in the fragrance compounds and fragrance raw materials used in cosmetic
products. The analysis is performed by gas chromatography and mass spectrometry (GC-MS) on matrix
samples which are “ready to be injected” and which are compatible with gas chromatography.
The analytes covered by this procedure are based on the contents of Tables 13.1 and 13.2 in the
SCCS 1459/11 opinion document (1) and as listed in the legislation proposed by the European
Commission. The rationale behind the final choice of procedure analytes is given in the table found in
Annex J.
The method was validated at IFRA and CEN level.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
— IEC Electropedia: available at http://www.electropedia.org/
— ISO Online browsing platform: available at https://www.iso.org/obp
4 Principle
This procedure has a calibration range between 2 ppm and 240 ppm, this permits the quantification of
suspected allergens in diluted matrices in the range 20 ppm to 2 400 ppm per analyte. Beyond the upper
concentration level the recommendation is that the sample should be diluted further, or that GC-FID (GC
with Flame Ionization Detection) is used preferably in combination with internal standard and response
factors.
The matrix samples are analysed for suspected allergens by GC-MS in a total of 4 runs, using 2 analyte
sets, both injected on two separate columns of differing polarity. Where necessary the matrix samples
should first be diluted in an appropriate solvent.
Their identification and quantification are achieved by selected ion monitoring (SIM; SIM-SCAN) mode
via the relative abundance of 3 characteristic fragment ions. The calculation and use of the corresponding
Q value or similar data evaluation factor can be applied and a ‘Decisional Tree’ (see Annex F, Figure F.1)
for the final inspection and validation of the data by a trained and experienced analyst is described. An
additional full-SCAN analysis is recommended to confirm the presence of the allergen in matrix samples
if only SIM methodology was used.
Their quantification is achieved in all modes by calibration using standard solutions and the internal
standards 1,4-dibromobenzene and 4,4’-dibromobiphenyl. The ‘Decisional Tree’ is employed to
determine the final concentration taking into account the different concentration values obtained from
analysis on both columns.
6

---------------------- Page: 8 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
5 Reagents
5.1 Solvents
5.1.1 Methyl pivalate
CAS [598-98-1]
Purity ≥ 99 %
Distillation to obtain a purity > 99,9 % is recommended to eliminate impurities, which are likely to
interfere with the signal of analytes such as terpenes. If a commercial grade is used then that shall be
analysed and confirmed as not giving rise to interfering artefacts.
5.1.2 Tert-Butyl Methyl Ether (MTBE)
CAS [1634-04-4]
Purity ≥ 99 %
Distillation to obtain a purity > 99,9 % is recommended to eliminate impurities, which are likely to
interfere with the signal of analytes such as terpenes. If a commercial grade is used then that shall be
analysed and confirmed as not giving rise to interfering artefacts.
Alternative solvents are available. If these are used for this method, then the Operator shall undertake
sufficient evaluation to verify that the solvent contains no constituents that would interfere with the
analytes in this method.
5.2 Reference samples (suspected allergens)
5.2.1 General
Initially, the purity of all standards should be measured by GC-FID (Annex A) if not certified by the
supplier. The precise CAS number of the target analyte and the rationale behind this is given in Annex J.
5.2.1.1 Acetyl cedrene alpha (main isomer in Vertofix® referred to in CAS number), CAS No.
[32388-55-9]
Highly variable composition with at least twelve constituents of molecular weight equal to 246 atomic
mass units (hereafter u).
5.2.1.2 Acetyl isoeugenol / Isoeugenyl acetate, CAS No. [93-29-8]
5.2.1.3 Amyl salicylate, pentyl salicylate CAS No. [2050-08-0]
This can contain isoamyl salicylate (CAS No. [87-20-7]), which should not be included in this assay.
5.2.1.4 Alpha amyl cinnamaldehyde (Flosal®), CAS No. [122-40-7, 78605-96-6]
2 possible isomers (E, Z); only E is quantified.
5.2.1.5 Alpha amylcinnamyl alcohol, CAS No. [101-85-9, 184900-07-0]
2 possible isomers (E, Z); only E is quantified.
5.2.1.6 Anethole, CAS No. [104-46-1, 4180-23-8]
2 possible isomers (E, Z); only E is quantified.
5.2.1.7 Anise alcohol, CAS No. [105-13-5]
7

---------------------- Page: 9 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
5.2.1.8 Benzaldehyde, CAS No. [100-52-7]
5.2.1.9 Benzyl alcohol, CAS No. [100-51-6]
5.2.1.10 Benzyl benzoate, CAS No. [120-51-4]
5.2.1.11 Benzyl cinnamate, CAS No. [103-41-3]
2 possible isomers (E,Z), only E is quantified.
5.2.1.12 Benzyl salicylate, CAS No. [118-58-1]
5.2.1.13 Camphor, CAS No. [76-22-2]
5.2.1.14 Carvone, CAS No. [99-49-0]
5.2.1.15 Caryophyllene beta, CAS No. [87-44-5]
5.2.1.16 Cinnamaldehyde, CAS No. [104-55-2, 14371-10-9]
2 possible isomers (E, Z), only E is quantified.
5.2.1.17 Cinnamyl alcohol, CAS No. [104-54-1, 4407-36-7]
2 possible isomers (E, Z), only E is quantified.
5.2.1.18 Citral, CAS No. [5392-40-5]
5.2.1.18.1 Neral
5.2.1.18.2 Geranial
Both Neral (Z isomer, CAS No. [106-26-3]) and Geranial (E isomer, CAS No. [141-27-5]) are quantified.
5.2.1.19 Citronellol, CAS No. [106-22-9]
5.2.1.20 Coumarin, CAS No. [91-64-5]
5.2.1.21 Damascenone beta (Rose Ketone-4), CAS No. [23696-85-7]
5.2.1.22 Damascone alpha, CAS No. [43052-87-5] covers both E and Z isomers
2 possible isomers (E, Z); only the E main isomer CAS No. [24720-09-0] (92 to 99) % is quantified.
5.2.1.23 Damascone beta (E), CAS No. [23726-91-2]
5.2.1.24 Damascone delta (Rose Ketone-3), CAS No. [57378-68-4]
3 possible isomers (trans/trans, cis/trans, trans/cis) only the major isomer (trans/trans, up to 90) % is
quantified.
5.2.1.25 Dimethylbenzylcarbinyl acetate (DMBCA), CAS No. [151-05-3]
5.2.1.26 Ebanol, CAS No. [67801-20-1]
5.2.1.26.1 Ebanol Isomer 1
8

---------------------- Page: 10 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
5.2.1.26.2 Ebanol Isomer 2
The 2 isomers (E, Z) are quantified.
5.2.1.27 Eugenol, CAS No. [97-53-0]
5.2.1.28 Eugenyl acetate, CAS No. [93-28-7]
5.2.1.29 Farnesol, CAS No. [4602-84-0]
Recommendation: Use E,E calibration curve to quantify other isomers.
4 possible isomers: (E,E) isomer is CAS No. [106-28-5] and the (Z,E) isomer is CAS No. [3790-71-4], and
two additional minor isomers.
5.2.1.30 Galaxolide (Hexamethylindanopyran), CAS No. [1222-05-5]
Only the two main isomers are quantified.
5.2.1.31 Geraniol, CAS No. [106-24-1]
5.2.1.32 Geranyl acetate, CAS No. [105-87-3]
5.2.1.33 Hexadecanolactone / Dihydroambrettolide, CAS No. [109-29-5]
5.2.1.34 Hexylcinnamaldehyde alpha, CAS No. [101-86-0, 165184-98-5]
2 isomers at least; only the E is quantified.
5.2.1.35 Hydroxycitronellal, CAS No. [107-75-5]
5.2.1.36 Tetramethylacetyloctahydronaphthalene (ISO E® Super)
NOTE This consists of several isomers including the alpha, (CAS No. [68155-66-8]); the major isomer, beta
(CAS No. [54464-57-2]); the gamma (CAS No. [68155-67-9]) and a minor (CAS No. [54464-59-4]).
5.2.1.37 Isoeugenol, CAS No. [97-54-1]
2 possible isomers (E, Z), only E- isoeugenol is quantified.
5.2.1.38 Isomethylionone alpha, CAS No. [127-51-5]
Can contain beta isomethylionone (CAS No. [79-89-0]), alpha methylionone (CAS No. [7779-30-8]), beta
methylionone (CAS No. [127-43-5]), pseudo isomethylionone (CAS No. [1117-41-5]).
Other grades may contain other isomers including the major one. These constituents should not be
assayed.
5.2.1.39 Butylphenyl methylpropional (Lilial®), CAS No. [80-54-6]
5.2.1.40 Limonene, CAS No. [138-86-3]
5.2.1.41 Linalool, CAS No. [78-70-6]
5.2.1.42 Linalyl acetate, CAS No. [115-95-7]
9

---------------------- Page: 11 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
5.2.1.43 Hydroxyisohexyl 3-cyclohexene carboxaldehyde (Lyral®), CAS No. [31906-04-4]
Contains Hydroxyisohexyl 4-cyclohexene carboxaldehyde (CAS No. 51414-25-6) (noted Lyral minor) that
has to be quantified.
5.2.1.44 Trimethyl-benzenepropanol (Majantol®), CAS No. [103694-68-4]
5.2.1.45 Menthol, CAS No. [1490-04-6, 89-78-1]
5.2.1.46 Methyl salicylate, CAS No. [119-36-8]
5.2.1.47 Methyl-2-octynoate (Folione®), CAS No. [111-12-6]
5.2.1.48 Pinene alpha, CAS No. [80-56-8]
5.2.1.49 Pinene beta, CAS No. [127-91-3]
5.2.1.50 3-Propylidene phthalide, CAS No. [17369-59-4]
The 2 isomers (E,Z) (E-isomer: CAS No. [56014-72-3] and Z-isomer: CAS No.[94704-89-9]) are quantified.
5.2.1.51 Salicylaldehyde, CAS No. [90-02-8]
5.2.1.52 Santalol, CAS No. [11031-45-1]
NOTE The 2 isomers, alpha-santalol (CAS No. [115-71-9]) and beta-santalol (CAS No. [77-42-9]), are quantified.
5.2.1.53 Sclareol, CAS No. [515-03-7]
5.2.1.54 Terpinene alpha, CAS No. [99-86-5]
5.2.1.55 Terpineol-alpha, CAS No. [98-55-5]
5.2.1.55.1 for terpineol gamma CAS No. [586-81-2]
5.2.1.55.2 for Terpineol cis-beta CAS No. [138-87-4]
5.2.1.55.3 for Terpineol trans-beta CAS No. [7299-41-4]
Recommendation: use alpha-terpineol calibration curve to quantify other isomers.
5.2.1.56 Terpinolene, CAS No. [586-62-9]
5.2.1.57 Vanillin, CAS No. [121-33-5]
5.3 Internal standards (ISTD)
5.3.1  1,4-Dibromobenzene (IS ), CAS No. [106-37-6], purity ≥ 98 %
A
5.3.2  4,4'-dibromobiphenyl (IS ), CAS No. [92-86-4], purity ≥ 98 %
B
10

---------------------- Page: 12 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
6 Apparatus
6.1 Gas chromatograph equipped with flame ionization detector (GC-FID)
This apparatus is only used to determine the purity of reference samples intended to be used for
calibration purposes and the ISTDs before quantitative analysis if required, or for the quantification of
high concentration analytes (above 2,4 % as found in essential oils for example). It is recommended to
perform the purity study using the procedure described in Annex A.
[3]
A GC-FID method is available for the measurement of analytes at concentrations above 2,4 %. Note
that this method does not implicitly cover all the allergens mentioned in this procedure: however, the
principle covered in that method can be applied to all the analytes contained within the proposed
legislation and this procedure.
6.2 Gas chromatograph coupled to a mass spectrometer (GC-MS)
6.2.1 General
This apparatus is used for quantitative analysis, to check for the presence and measure the concentration
of the suspected allergens. The system shall be able to comply with the following recommendation.
6.2.2 GC-MS system
a) Equipped with electronic control of carrier gas pressures and/or flows;
b) it is recommended that an autosampler, fitted with a syringe of suitable size, be used for the injection
of the calibration and sample solutions;
c) the glass injection liner shall be inert with an interior volume compatible with the expansion volume
of the dilution solvent;
d) two capillary columns of different phase types are to be used for quantitative analysis;
e) mass spectrometer tuning and the levels of air and water should be checked regularly to ensure
optimal sensitivity, as well as the injector cleanliness to maintain analytical performance.
The sensitivity of the mass spectrometer should also be optimized through maximizing the detector
signal to noise ratio. One option could be to increase the detector (electron multiplier) voltage according
to the instrument capabilities and manufacturers recommendations.
6.3 Capillary columns for GC
Whilst the retention times and SIM windows have been specified for the columns mentioned, these are
for guidance only and the user shall verify all compound retention times and the associated SIM windows
for their own installations.
The column lengths are those used for the initial development and validation by the AWG. In practice,
additional separation of the target analyte may be achieved by moving to a longer column (50 m or 60 m).
If such columns are used, then it is the user’s responsibility to validate the performance of that column
type/length in the context of this method (resolution; SIM windows) and validate that length for the
analytes(s) being investigated (see Annex B).
11

---------------------- Page: 13 ----------------------
SIST EN 16274:2022
EN 16274:2021 (E)
Table 1 — Recommended GC Column Phase/Types and Analytical Parameters
Dimensions Carrier MS Mean
Recommended
(Length × gas conditions resolution
oven Injection
Phase internal

temperature conditions R
diameter × thic
program
kness)
Helium Source SIM method
Constant Temperature 1
80 °C for 4 min,
flow at at 250 °C –
1.43
then 15 °C/min
1,2 mL/m Quadrupole
to 105 °C for 2
in temp. at
100 % min,
(velocity 150 °C –
30 m × 0,25 mm
Split/
Polydimethylsiloxane then 4 °C/min to
approx. Scanning
× 0,25 µm
SIM method
(PDMS) 150 °C, then splitless
40 cm/se mass range
2
10 °C/min to injector at
c) (m/z) from
280 °C – hold 5 250 °C -
1.40
35 Da to 350
min. Deactivated
Da
liner-
-
Inject 1 µL
Ionization
SIM method
with split
energy:
1:20 – Inj. 1
70 eV
Temp.
1.40
80 °C for 1 min,
250 °C
then 10
...

SLOVENSKI STANDARD
oSIST prEN 16274:2020
01-marec-2020
Analizne metode za alergene - Kvantitativno določevanje 57 domnevnih alergenov
z razširjenega seznama v dišavnih izdelkih s plinsko kromatografsko analizo
vzorcev, ki so pripravljeni za injeciranje
Method for Analysis of Allergens - Quantification of an extended list of 57 suspected
allergens in ready to inject fragrance materials by gas chromatography mass
spectrometry
Analyseverfahren für Allergene - Quantifizierung einer erweiterten Liste von zu
vermutenden Allergenen in einspritzfertigen Duftstoffen mittels
Gaschromatographie/Massenspektrometrie
Méthodes d'analyse des allergènes - Quantification des fragrances allergènes
suspectées dans les produits de consommation - Étape 1 : Analyse par GC
d'échantillons prêts à être injectés
Ta slovenski standard je istoveten z: prEN 16274
ICS:
71.040.50 Fizikalnokemijske analitske Physicochemical methods of
metode analysis
71.100.70 Kozmetika. Toaletni Cosmetics. Toiletries
pripomočki
oSIST prEN 16274:2020 en,fr,de
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.

---------------------- Page: 1 ----------------------
oSIST prEN 16274:2020

---------------------- Page: 2 ----------------------
oSIST prEN 16274:2020


DRAFT
EUROPEAN STANDARD
prEN 16274
NORME EUROPÉENNE

EUROPÄISCHE NORM

February 2020
ICS 71.100.60 Will supersede EN 16274:2012
English Version

Method for Analysis of Allergens - Quantification of an
extended list of 57 suspected allergens in ready to inject
fragrance materials by gas chromatography mass
spectrometry
Méthodes d'analyse des allergènes - Quantification des Analyseverfahren für Allergene - Quantifizierung einer
fragrances allergènes suspectées dans les produits de erweiterten Liste von zu vermutenden Allergenen in
consommation - Étape 1 : Analyse par GC einspritzfertigen Duftstoffen mittels
d'échantillons prêts à être injectés Gaschromatographie/Massenspektrometrie
This draft European Standard is submitted to CEN members for enquiry. It has been drawn up by the Technical Committee
CEN/TC 347.

If this draft becomes a European Standard, CEN members are bound to comply with the CEN/CENELEC Internal Regulations
which stipulate the conditions for giving this European Standard the status of a national standard without any alteration.

This draft European Standard was established by CEN in three official versions (English, French, German). A version in any other
language made by translation under the responsibility of a CEN member into its own language and notified to the CEN-CENELEC
Management Centre has the same status as the official versions.

CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia,
Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway,
Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey and
United Kingdom.

Recipients of this draft are invited to submit, with their comments, notification of any relevant patent rights of which they are
aware and to provide supporting documentation.

Warning : This document is not a European Standard. It is distributed for review and comments. It is subject to change without
notice and shall not be referred to as a European Standard.


EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION

EUROPÄISCHES KOMITEE FÜR NORMUNG

CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2020 CEN All rights of exploitation in any form and by any means reserved Ref. No. prEN 16274:2020 E
worldwide for CEN national Members.

---------------------- Page: 3 ----------------------
oSIST prEN 16274:2020
prEN 16724:2020 (E)
Contents Page
European foreword . 4
Introduction . 5
1 Scope . 6
2 Normative references . 6
3 Terms and definitions . 6
4 Principle . 6
5 Reagents . 7
5.1 Solvents . 7
5.2 Reference samples (suspected allergens) . 7
5.3 Internal standards (ISTD) . 11
6 Apparatus . 11
6.1 Gas chromatograph equipped with flame ionization detector (GC-FID) . 11
6.2 Gas chromatograph coupled to a mass spectrometer (GC-MS) . 11
6.3 Capillary columns for GC . 11
6.4 Analytical Balance . 12
7 Mass Spectrometer Acquisition Conditions . 12
7.1 Establishment of Retention Times and SIM chromatograph Windows . 12
7.2 SIM Window – Set up and Criteria . 13
7.3 GC-MS Scan Mode Verification . 13
8 Stock and Sample solutions - preparation and storage . 13
8.1 General information . 13
8.2 Preparation of stock solutions from reference samples . 14
8.3 Preparation of internal standard solutions and calibration solutions . 14
9 Sample Analysis . 17
9.1 General. 17
9.2 Sample Preparation for Analysis. 17
9.3 Sequence . 18
10 Data validation and treatment . 19
10.1 General. 19
10.2 Examination of Q Values . 19
10.3 Relative intensity of characteristic ions in SIM . 20
10.4 Relative ion intensity using scan data . 20
10.5 Data verification scheme and reporting of final concentration . 20
10.6 GC-MS scan mode verification . 20
10.7 Mass spectrum evaluation in SCAN mode . 20
11 Assay report . 21
12 References: . 21
Annex A (informative) Determination of calibrant and reference sample purity . 22
Annex B (informative) GC capillary column parameters. 28
Annex C (informative) SIM Ions for SIM or SIM-SCAN Use . 29
2

---------------------- Page: 4 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)
Annex D (informative) Example of SIM Windows . 38
Annex E (informative)  Example of chromatograms . 46
Annex F (normative) Decisional tree for quantification of suspected allergens . 48
Annex G (informative)  Preparation of stock solutions from reference samples . 49
Annex H (informative)  Calibration methods and approach . 54
Annex I (informative)  Quantification of allergens . 56
Annex J (informative)  Suspected allergens selected as analytes targets – Rationale . 59
Annex K (informative)  Other general information . 69

Tables
Table 1 — Recommended GC Column Phase/Types & Analytical Parameters . 12
Table 2 — Preparation volume for Solution A3 . 15
Table 3 — Preparation volume for Solution A4 . 15
Table 4 — Individual Allergen Calibration Standard Solution concentrations . 17
Table 5 — Maximum allowable tolerances for the relative ion intensities (expressed in percentage of
intensity of the most abundant ion i.e. the base peak) . 20
Table C.1 — SIM ions (m/z) of suspected allergens listed in the elution order on 100%PDMS-type
column, corresponding SIM methods (1 or 2) and internal standard (ISA or ISB) used for quantification
. 29
Table C.2 — SIM ions (m/z) of allergens listed in the elution order on 50%Polydimethylsiloxane
50%Phenylmethylpolysiloxane -type column, corresponding SIM methods (1 or 2) and internal
standard (ISA or ISB) used for quantification . 32
Table C.3 — Alternative SIM ions (m/z) of allergens listed in the elution order on 100% PDMS or 00%
PDMS type columns, corresponding SIM methods (1 or 2) and internal standard (ISA or ISB) used for
quantification . 35
Table D.1 — Example of windows for SIM1 method on 100% PDMS type-column . 38
Table D.2 — Example of windows for SIM2 method on 100% PDMS type-column . 40
Table D.3 — Example of windows for SIM1 method on 50%Polydimethylsiloxane
50%Phenylmethylpolysiloxane type-column . 42
Table D.4 — Example of windows for SIM2 method on 50%Polydimethylsiloxane
50%Phenylmethylpolysiloxane type-column . 44
Table G.1 — Required mass for the preparation of Solution A1 . 51
Table G.2 — Required mass for the preparation of Solution A2 . 52
Table I.1 — Resolution of ∆ for the determination of C . 58
0
Table J.1 — Allergens Selected as Analysis Targets – Rationale . 59

3

---------------------- Page: 5 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)

European foreword
This document (prEN 16274:2020) has been prepared by Technical Committee CEN/TC 347 “Methods
for analysis of allergens”, the secretariat of which is held by SNV.
This document is currently submitted to the CEN Enquiry.
This document will supersede EN 16274:2012.
This document has been prepared under a mandate given to CEN by the European Commission and the
European Free Trade Association, and supports essential requirements of EU Directive(s).
4

---------------------- Page: 6 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)
Introduction
Directive 2003/15/EC amending Council Directive 76/768/EEC, relating to cosmetic products, regulates
the obligation to inform consumers of the presence of 24 chemically-defined fragrance substances
identified as potential allergens in cosmetic products. Following the publication of the Scientific
Committee on Consumer Safety’s document (SCCS/1459/11), it was proposed to extend that to 57
fragrance substances, some of them existing under several isomeric forms or as mixtures. This required
the development of a new quantification method in response to the evolution of regulatory requirements.
The new analytical method has been developed using gas chromatography and mass spectrometry (GC-
MS), to detect and to quantify the 57 fragrance substances and their relevant isomers in ready to inject
fragrance raw materials and oils and dilutions thereof.
The method described in this document does not include requirements for the preparation of samples in
matrices for which direct injection in GC is not feasible.
The present document describes a working analytical method based on IFRA Analytical Working Group
developments. The analytical method was validated based on a ring test performed by the CEN working
group using the accuracy profile approach.
5

---------------------- Page: 7 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)

1 Scope
The present method permits the identification and quantification of the volatile compounds suspected as
allergens, which are present in the fragrance compounds and raw materials used in cosmetic products.
The analysis is performed by gas chromatography and mass spectrometry (GC-MS) on matrix samples
which are “ready to be injected” and which are compatible with gas chromatography.
The analytes covered by this procedure are based on the contents of Tables 13.1 and 13.2 in the SCCS
1459/11 opinion document (1) and as listed in the legislation proposed by the European Commission.
The rationale behind the final choice of procedure analytes is given in the table found in Annex J.
The method was validated at IFRA and CEN level with two solvents, namely methyl pivalate and tert-
methyl butyl ether.
Many of the Allergens in the SCCS ‘Opinion’ documentation submitted to the EU Commission consist of
multiple isomers of the same molecules and/or complex mixtures.
In many cases, this complexity can be resolved by this Procedure but that leaves a number of target
analytes where this isn’t the case.
In addition, the SCCS Opinion also documents a number of CAS Numbers related to the same parent
molecule. These are related to the optically active forms of the similar parent structure. For this
procedure, the optically active forms are not considered as these require special (and in some cases, very
different) GC phases to achieve the required separation.
2 Normative references
There are no normative references in this document.
3 Terms and definitions
No terms and definitions are listed in this document.
ISO and IEC maintain terminological databases for use in standardization at the following addresses:
• IEC Electropedia: available at http://www.electropedia.org/
• ISO Online browsing platform: available at http://www.iso.org/obp
4 Principle
This procedure has a calibration range between 2 ppm and 240 ppm, this permits the quantification of
suspected allergens in diluted matrices in the range 20 ppm to 2 400 ppm per analyte; beyond the upper
concentration level the recommendation is that the sample should be diluted further, or that GC-FID (GC
with Flame Ionisation Detection) is used preferably in combination with internal standard and response
factors.
The matrix samples are analysed for suspected allergens by GC-MS in a total of 4 runs, using 2 analyte
sets, both injected on two separate columns of differing polarity. Where necessary the matrix samples
should first be diluted in an appropriate solvent.
Their identification and quantification are achieved by selected ion monitoring (SIM; SIM-SCAN) mode
via the relative abundance of 3 characteristic fragment ions. The calculation and use of the corresponding
Q value or similar data evaluation factor can be applied and a ‘Decisional Tree’ (see Annex F) for the final
inspection and validation of the data by a trained and experienced analyst is described. An additional full-
SCAN analysis is recommended to confirm the presence of the allergen in matrix samples if only SIM
methodology was used.
6

---------------------- Page: 8 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)
Their quantification is achieved in all modes by calibration using standard solutions and the internal
standards 1,4-dibromobenzene and 4,4-dibromobiphenyl. The ‘Decisional Tree’ is employed to
determine the final concentration taking into account the different concentration values obtained from
analysis on both columns.
This Procedure is written as a methodology to be applied in a working laboratory (QC and R&D type).
Particular attention shall be paid by the user to the details contained in the Annexes and to specific
comments (‘Notes’) in the text regarding the application of this methodology.
5 Reagents
5.1 Solvents
5.1.1 Methyl pivalate
CAS [598-98-1]
Purity ≥ 99%
Distillation to obtain a purity >99.9 % is recommended to eliminate impurities, which are likely to
interfere with the signal of analytes such as terpenes. If a commercial grade is used then that shall be
analysed and confirmed as not giving rise to interfering artefacts.
5.1.2 Tert-Butyl Methyl Ether (MTBE)
CAS [1634-04-4]
Purity ≥ 99%
Distillation to obtain a purity >99.9 % is recommended to eliminate impurities, which are likely to
interfere with the signal of analytes such as terpenes. If a commercial grade is used then that shall be
analysed and confirmed as not giving rise to interfering artefacts.
Alternative solvents are available. If these are used for this method, then the Operator shall undertake
sufficient evaluation to verify that the solvent contains no constituents that would interfere with the
analytes in this method.
5.2 Reference samples (suspected allergens)
Initially, the purity of all standards should be measured by GC-FID (Annex A) if not certified by the
supplier. The precise CAS number of the target analyte and the rationale behind this is given in Annex J.
5.2.1 Acetyl cedrene alpha (main isomer in Vertofix® referred to in CAS number), CAS No.
[32388-55-9]
NOTE Highly variable composition with at least twelve constituents of molecular weight equal to 246 amu.
5.2.2 Acetyl isoeugenol / Isoeugenyl acetate, CAS No. [93-29-8]
5.2.3 Amyl salicylate, pentyl salicylate CAS No. [2050-08-0]
This can contain isoamyl salicylate (CAS No. [87-20-7]), which should not be included in this assay.
5.2.4 Alpha amyl cinnamaldehyde (Flosal®), CAS No. [122-40-7, 78605-96-6]
NOTE 2 possible isomers (E, Z); only E is quantified.
7

---------------------- Page: 9 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)

5.2.5 Alpha amylcinnamyl alcohol, CAS No. [101-85-9, 184900-07-0]
NOTE 2 possible isomers (E, Z); only E is quantified.
5.2.6 Anethole, CAS No. [104-46-1, 4180-23-8]
NOTE 2 possible isomers (E, Z); only E is quantified.
5.2.7 Anise alcohol, CAS No. [105-13-5]
5.2.8 Benzaldehyde, CAS No. [100-52-7]
5.2.9 Benzyl alcohol, CAS No. [100-51-6]
5.2.10 Benzyl benzoate, CAS No. [120-51-4]
5.2.11 Benzyl cinnamate, CAS No. [103-41-3];
NOTE Only E is quantified.
5.2.12 Benzyl salicylate, CAS No. [118-58-1]
5.2.13 Camphor, CAS No. [76-22-2]
5.2.14 Carvone, CAS No. [99-49-0]
5.2.15 Caryophyllene beta, CAS No. [87-44-5]
5.2.16 Cinnamaldehyde, CAS No. [104-55-2, 14371-10-9]
NOTE 2 possible isomers (E, Z), only E is quantified.
5.2.17 Cinnamyl alcohol, CAS No. [104-54-1, 4407-36-7]
NOTE 2 possible isomers (E, Z), only E is quantified.
5.2.18 Citral, CAS No. [5392-40-5]
NOTE Both Neral (Z isomer, CAS No. [106-26-3]) and Geranial (E isomer, CAS No. [141-27-5]) are quantified.
5.2.19 Citronellol, CAS No. [106-22-9]
5.2.20 Coumarin, CAS No. [91-64-5]
5.2.21 Damascenone beta (Rose Ketone-4), n° CAS [23696-85-7]
5.2.22 Damascone alpha, CAS No. [43052-87-5]
NOTE 2 possible isomers (E, Z); only the E main isomer (92 to 99%) is quantified.
5.2.23 Damascone beta (E), CAS No. [23726-91-2]
5.2.24 Damascone delta (Rose Ketone-3), CAS No. [57378-68-4]
NOTE 3 possible isomers (trans/trans, cis/trans, trans/cis) only the major isomer (trans/trans, up to 90%) is
quantified.
8

---------------------- Page: 10 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)
5.2.25 Dimethylbenzylcarbinyl acetate (DMBCA), CAS No. [151-05-3]
5.2.26 Ebanol, CAS No. [67801-20-1]
NOTE The 2 isomers (E, Z) are quantified.
5.2.27 Eugenol, CAS No. [97-53-0]
5.2.28 Eugenyl acetate, CAS No. [93-28-7]
5.2.29 Farnesol, CAS No. [4602-84-0]
NOTE 4 possible isomers: (E,E) isomer is (n° CAS [106-28-5]) and the (Z,E) isomer is (n° CAS [3790-71-4]).
Recommendation: Use E,E calibration curve to quantify other isomers.
5.2.30 Galaxolide (Hexamethylindanopyran), CAS No. [1222-05-5]
NOTE Only the two main isomers are quantified.
5.2.31 Geraniol, CAS No. [106-24-1]
5.2.32 Geranyl acetate, CAS No. [105-87-3]
5.2.33 Hexadecanolactone / Dihydroambrettolide, CAS No. [109-29-5]
5.2.34 Hexylcinnamaldehyde alpha, CAS No. [101-86-0, 165184-98-5]
NOTE 2 isomers at least; only the E is quantified.
5.2.35 Hydroxycitronellal, CAS No. [107-75-5]
5.2.36 Tetramethylacetyloctahydronaphthalene (ISO E® Super)
NOTE This consists of several isomers including the alpha, (CAS No. [68155-66-8]); the major isomer, beta
(CAS No. [54464-57-2]); the gamma (CAS No. [68155-67-9]) and a minor (CAS No. [54464-59-4]).
5.2.37 Isoeugenol, CAS No. [97-54-1]
NOTE 2 possible isomers (E, Z), only E- iso eugenol is quantified.
5.2.38 Isomethylionone alpha, CAS No. [127-51-5]
NOTE Can contain beta isomethylionone (CAS No. [79-89-0]), alpha methylionone (CAS No. [7779-30-8]), beta
methylionone (CAS No. [127-43-5]), pseudo isomethylionone (CAS No. [1117-41-5]).
Other grades may contain other isomers including the major. These constituents should not be assayed.
9

---------------------- Page: 11 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)

5.2.39 Butylphenyl methylpropional (Lilial®), CAS No. [80-54-6]
5.2.40 Limonene, CAS No. [138-86-3]
5.2.41 Linalool, CAS No. [78-70-6]
5.2.42 Linalyl acetate, CAS No. [115-95-7]
5.2.43 Hydroxyisohexyl 3-cyclohexene carboxaldehyde (Lyral®), CAS No. [31906-04-4]
NOTE Contains Hydroxyisohexyl 4-cyclohexene carboxaldehyde (CAS No.51414-25-6) (noted Lyral minor),
has to be quantified.
5.2.44 Trimethyl-benzenepropanol (Majantol®), CAS No. [103694-68-4]
5.2.45 Menthol, CAS No. [1490-04-6, 89-78-1]
5.2.46 Methyl salicylate, CAS No. [119-36-8]
5.2.47 Methyl-2-octynoate (Folione®), CAS No. [111-12-6]
5.2.48 Pinene alpha, CAS No. [80-56-8]
5.2.49 Pinene beta, CAS No. [127-91-3]
5.2.50 3-Propylidene phthalide, CAS No. [17369-59-4]
NOTE The 2 isomers (E,Z) (E-isomer: CAS No.56014-72-3 and Z- isomer: CAS No.94704-89-9) are quantified
(isomers ratio near 1:6).
5.2.51 Salicylaldehyde, CAS No. [90-02-8]
5.2.52 Santalol CAS No. [11031-45-1]
NOTE The 2 isomers, alpha-santalol (CAS No. [115-71-9]) and beta-santalol (CAS No. [77-42-9]), are quantified.
5.2.53 Sclareol, CAS No. [515-03-7]
5.2.54 Terpinene alpha, CAS No. [99-86-5]
5.2.55 Terpineol-alpha, CAS No. [98-55-5] for alpha-terpineol; 5.2.55a for gamma-terpineol CAS
No. [586-81-2]; 5.2.55b for Terpineol cis-beta CAS No. [138-87-4]; 5.2.55c for Terpineol trans-
beta CAS No. [7299-41-4]
Recommendation: use alpha-terpineol calibration curve to quantify other isomers.
10

---------------------- Page: 12 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)
5.2.56 Terpinolene, CAS No. [586-62-9]
5.2.57 Vanillin, CAS No. [121-33-5]
5.3 Internal standards (ISTD)
5.3.1 1,4-Dibromobenzene (ISA), CAS No. [106-37-6], purity ≥ 98%.
5.3.2 4,4'-dibromobiphenyl (IS ), v [92-86-4], purity ≥ 98%.
B
6 Apparatus
6.1 Gas chromatograph equipped with flame ionization detector (GC-FID)
This apparatus is only used to determine the purity of reference samples intended to be used for
calibration purposes and the ISTDs before quantitative analysis if required, or for the quantification of
high concentration analytes (>2.4% as found in essential oils for example). It is recommended to perform
the purity study using the procedure described in Annex A.
[3]
A GC-FID method is available for the measurement of analytes at concentrations >2.4%. Note that this
method does not implicitly cover all the allergens mentioned in this procedure: however, the principle
covered in that method can be applied to all the analytes contained within the proposed legislation and
this procedure.
6.2 Gas chromatograph coupled to a mass spectrometer (GC-MS)
6.2.1 General
This apparatus is used for quantitative analysis, to check for the presence and measure the concentration
of the suspected allergens. The system shall be able to comply with the following requirements:
6.2.2 GC-MS System
a. Equipped with electronic control of carrier gas pressures and/or flows.
b. It is recommended that an autosampler, fitted with a syringe of suitable size, be used for the injection
of the calibration and sample solutions.
c. The glass injection liner shall be inert with an interior volume compatible with the expansion volume
of the dilution solvent.
d. Two capillary columns of different phase types are to be used for quantitative analysis.
e. Mass spectrometer tuning and the levels of air and water should be checked weekly, as well as the
injector cleanliness to maintain analytical performance.
The sensitivity of the mass spectrometer should also be optimized through maximising the detector
signal to noise ratio. One option could be to increase the detector (electron multiplier) voltage according
to the instrument capabilities and manufacturers recommendations.
6.3 Capillary columns for GC
Whilst the retention times and SIM windows have been specified for the columns mentioned, these are
for guidance only and the user shall verify all compound retention times and the associated SIM windows
for their own installations.
11

---------------------- Page: 13 ----------------------
oSIST prEN 16274:2020
prEN 16274:2020 (E)

NOTE The column lengths are those used for the initial development and validation by the AWG. In practice,
additional separation of the target analyte may be achieved by moving to a longer column (50 or 60m). If such
columns are used, then it is the user’s responsibility to validate the performance of that column type/length in the
context of this me
...

Questions, Comments and Discussion

Ask us and Technical Secretary will try to provide an answer. You can facilitate discussion about the standard in here.